FCR-7-xx-400-2-45-IND - Reflection Probe with Changeable Tip for Powdery Substances | |
|
|
The reflection probe FCR-7xx-400-2-45-IND with changeable tip is excellent for measuring absorption and reflection in liquids and powdery substances. During measurement simply dip the waterproof probe tip in the medium to be examined. The probe end consists of a saphirre window glued in the tip. The light led into the tip over 6 optical fibers is directed to the sample at a 45° angle. Insofern This is important because if the angle is strictly followed direct reflections of the windows can be avoided. The material to be analyzed reflects the incident light which is directed back to the spectrometer over a 7th optical fiber. Both fiber bundles protected by a flexible metal protection tube are connected over an SMA-connector.
Specifications | FCR-7-xx-400-2-45-IND |
Spectral range | 200 - 800 nm (UV/VIS)
350 - 1100 nm (NIR) |
Measuring head | Stainless steel cylinder, waterproof,
Length: 150 mm, Outer diameter: 13 mm (1/2") |
Probe material | Stainless steel, fused silica window
Diameter: 10 mm, thickness: 1mm |
Optical fibers | 7 x 400µm glass fibers,
Numerical aperture: 0,22,
6 illumination fibers, 1 receiving fiber |
Fiberoptic connectors | 2 x SMA |
Tubing | Inner: teflon-tube,
Outer: flexible chrome plated brass with stress relieve,
Diameter: 5 mm |
Temperature range | - 40° C - 100° C (other on request) |
Minimal bend radius | Short term: 20mm, long term: 60mm |
|
|
|
Last change 06/12/2007 11:48 AM