SentroProc NIR QC - NIR-Spectrometer for quality control



Request for information





SentroProc NIR QC was developed for routine analysis in the industrial quality laboratory. Due to its considerable lower price in relation to the already well known NIR-Systems, established NIR-analysis can be made less cost intensive and new kinds of applications can be started. The robust system uses a NIR diode array spectrometer with 256 or 512 elements. The system can be expanded to the visible region. Probe heads in different versions for measurement in transmission and / or reflection complete the system. Two different measurement heads can be installed in simultaneously, swithcing the measurement between them is simple.
An additional PC is not necessary. All essential software packages are installed on the integrated embedded PC running with Win2000 ready for operation. USB- and Ethernet-Interfaces make it possible to connect external devices and transmitting data in a network.
By using chemometric models, calibrations for chemical and physical parameters can be applied.
SentroProc NIR QC is been used for qualitative measurements and identification of materials.


Applications of the SentroProc NIR QC
    • Measurement of the OH- and Acid-number
    • Identification of materials
    • Quality control
    • Constituents of minerals and soil
    • Identification of paper and pulp
    • Detection and measuring of thin films

Quantitative analysis
Calibrations for chemical and physical parameters by using chemometric models can be created. There must be calibration samples measured with a reference method. The numbers of samples can be twenty to some hundreds depending on the robustness and the number of parameters you want to determine.


Features
  • Determination of up to 10 parameters at once
  • Comfortable generating of models using GRAMS-Software (Thermo Galactic), have to be ordered separately
  • Easy choice of the products that wanted to measured out of a data source of products


Qualitative measurements and identification
SentroProc NIR QC can also be used for qualitative measurements and material identification. For this the user can build up a data base of spectra that will be used later to identify materials automatically. Documentation and archiving can be done in the data base too.


Measurement procedure
Inhomogeneous sample can be measured with an easy to use measuring head. The distance between head and sample is approx. 50 mm. The sample can be rotating in a petri dish under the head and can be measured non tactile. Fluids can be measured in transmission using cuvette holder for cuvettes 1mm to 10 mm optical path. The cuvettes can be temperature controlled optionally. Beside fluids glasses, filters and foils can also be measured in transmission.

Technical specification
Spectral range 900 – 1680nm, expandable to 400 – 1680nm
Measurement principleDiffuse Reflection (measurement heads for different types of samples)
Transmission (in cuvettes or with dip probes)
Spectral resolutionca. 8nm FWHM (256 Pixel, 3.2nm/Pixel)
Measuring time Transmission: < 5s ( 500 – 1000 means)
Reflection: < 30 – 60s (150 – 300 means)
Photometric stabilityTransmission: < 0.2 mAU
Reflection: < 1mAU (spectral range 1000 – 1650nm)
Reference intervallapprox. 1 hour
InterfacesUSB and Ethernet
Light sourceTungsten-Halogene-Lightsource with 10000h liftetime
2850 K color temperature, Power: 5 W, Warm up timet: 1h
OptionsTurn table for petri dishes up to 100mm diameter ( 50 mm distance of the sample)
Transmissions dip probe
System for transmission and reflection (2 measuring heads)
System with temperature regulates sample holder


Software WinSpec Proc
    • Quantitative determination using chemometric models created by the software GRAMS
    • Identification
    • Easy to use
    • Internal log-files and protocol
    • Internal system checks
    • Different user levels for measurement, model administration and setups
    • Data can be stored as files and in a data base
    • Link to ACCESS-database


Last change 08/14/2007 03:47 PM